sarbi engineering
 
  profile
  test equipments
  Blood Bank Equipment
  Pharma Lab Equipment
  test specimens
  contact us
  home  
your total lubrication solution
  sarbi petroleum sarbi engineering
 
Test Equipment  
   
 
  AUTOMATIC FREEZING POINT APPARATUS FOR ENGINE COOLANT
  STANDARDS: JIS K 2234
   
 

This method covers the determination of the freezing point of an aqueous engine coolant solution in the laboratory by determination of the time temperature curve.

Features:

 

PC acquires data on line and displays. The apparatus automatically plots the line temperature curve from which freezing point is deduced.

   
 
   
  Apparatus Consists of:
   
 
-60°C Cryogenic circulator bath
Cooling bath Houses cooling tube
Cooling Tube Jacketed with polymer stopper
Motorized Stirring The stirrer rod is coil type made of steel wire 1.6mm in diameter. The stirrer rod moves up & down by connected motor assembly
Digital Thermometer & Sensor Range is from -50° to 50°C. The temperature is recorded every 1 minute initially and then every 15 seconds to the nearest 0.1° C.
Data logger & Software For logging time v/s temperature relation. The Software gives direct plotted graph wherein one can determine the freezing point.